5-268 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:CONStant:VSU<n>[:SOURce]
:PAGE:STRess:SETup:CONStant:VSU<n>[:SOURce]
This command sets the constant SOURCE value of VSU<n> for stress.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
The specified VSU must not be DISable.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:CONStant:VSU<n> [ :SOURce] source | MINimum |
MAXimum
Parameter
Query response source <newline><^END>
source is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:CONS:VSU1 10"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:VSU1 MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:VSU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:VSU1? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:DURation
This command sets the stress time in seconds.
If :SETup:MODE is PCOunt, this parameter is not used.
At *RST, this value is 1 ms.
Syntax :PAGE:STRess:SETup:DURation duration | MINimum | MAXimum
Parameter Type Explanation
source numeric -20 to 20 V

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals