Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-197
SCPI Commands
:PAGE:MEASure:QSCV:LNAMe
:PAGE:MEASure:QSCV:LNAMe
This command sets the variable name of the leakage current data measured during
the Quasi-static CV measurements. This command also has a query form.
At *RST, this value is set to LEAK.
Syntax :PAGE:MEASure:QSCV:LNAMe name
:PAGE:MEASure:QSCV:LNAMe?
Parameter
Query response name <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:LNAME 'ILEAK'"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:LNAME?"
ENTER @Hp4155;A$
:PAGE:MEASure:QSCV[:MENU]
This command displays the MEASURE: QSCV SETUP screen.
Available only in the Quasi-static CV measurement mode. An error occurs if this
command is entered in another measurement mode.
Syntax :PAGE:MEASure:QSCV[:MENU]
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV"
See also :PAGE:CHANnels:CDEFinition:MODE
Parameter Type Explanation
name string String of up to 6 alphanumeric characters.
The first character must be an alpha character.