Reference 19
InfiniiVision 7000B Series Oscilloscopes User’s Guide 417
X Units A 32- bit integer that identifies the unit of measure for X values
in the acquired data:
0 = Unknown.
1 = Volts.
2 = Seconds.
3 = Constant.
4 = Amps.
5 = dB.
6 = Hz.
Y Units A 32- bit integer that identifies the unit of measure for Y values
in the acquired data. The possible values are listed above under “X Units”.
Date A 16- byte character array, left blank in InfiniiVision oscilloscopes.
Time A 16- byte character array, left blank in the InfiniiVision
oscilloscopes.
Frame A 24 byte character array that is the model number and serial
number of the oscilloscope in the format of: MODEL#:SERIAL#.
Waveform Label A 16 byte character array that contains the label assigned
to the waveform.
Time Tags A 64- bit double, only used when saving multiple segments
(requires segmented memory option). This is the time (in seconds) since
the first trigger.
Segment Index A 32- bit unsigned integer. This is the segment number.
Only used when saving multiple segments.
Waveform Data Header
A waveform may have more than one data set. Each waveform data set
will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals