4 Triggering
130 InfiniiVision 7000B Series Oscilloscopes User’s Guide
< qualifier time set softkey
When the less than (<) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width less than the time value
displayed on the softkey.
When the time range (><) qualifier is selected, the Entry knob sets the
upper time range value.
> qualifier time set softkey
When the greater than (>) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width greater than the time value
displayed on the softkey.
When the time range (><) qualifier is selected, the Entry knob sets the
lower time range value.