7 Acquisition Modes
234 InfiniiVision 7000B Series Oscilloscopes User’s Guide
Segmented Memory
You can purchase the oscilloscope with the segmented memory option
factory-installed (Option SGM) or you can easily install it yourself (order
model number N5454A “Segmented Memory”).
When capturing multiple infrequent trigger events it is advantageous to
divide the oscilloscope’s memory into segments. This lets you capture
signal activity without capturing long periods of signal “dead time.”
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using Segmented memory, use the Analyze Segments feature (see
“Measurements, statistics, and infinite persistence” on page 236) to show
infinite persistence across all acquired segments. See also “To set or clear
infinite persistence” on page 85 for details.
Using segmented memory
1 Set up a trigger condition. (See “Triggering” on page 95 for details.)
2 Press the [Acquire] key in the Waveform section of the front panel.
3 Press the Segmented softkey to access the Segmented Memory Menu.
4 In the Segmented Memory Menu, press the Segmented softkey to enable
segmented memory acquisitions.
5 Press the #ofSegs softkey and rotate the Entry knob to select the
number of segments into which you would like to divide the
oscilloscope’s memory. Memory can be divided into as few as two
segments and as many as 2000 segments.
6 Press the [Run] key or the [Single] key.
The oscilloscope will run and fill a memory segment for each trigger
event. When the oscilloscope is busy acquiring multiple segments, the
progress is displayed in the upper right area of the display. The
oscilloscope will continue to trigger until memory is filled, then the
oscilloscope will stop.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals