Displaying Waveforms 3
InfiniiVision 7000B Series Oscilloscopes User’s Guide 69
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y
axis. Again, note the Y value.
Figure 5 Cursors set to center of signal
7 Calculate the phase difference using the formula below.
= sin
second Y
first Y
--------------------------
1.031
1.688
-------------
; = 37.65 degrees of phase shift=