Triggering 4
InfiniiVision 7000B Series Oscilloscopes User’s Guide 137
7 If you select the Timeout term, you will need to set a timeout value.
a Press the Ter m softkey and select Timeout.
b Press the Timeout softkey, then turn the Entry knob to set the
timeout value.
Timeout can be set from 10 ns to 10 seconds. The timer starts when
the Find condition has been met. If the timer is running and another
Find condition occurs, the timer will restart from time zero.
Adjust the Trigger Level
• For analog channels, adjust the trigger level for the selected analog
channel by turning the Trigger Level knob.
• To set the threshold level for digital channels, press the [Digital] key and
select Thresholds.
The value of the trigger level or digital threshold is displayed in the
upper- right corner of the display.