EasyManua.ls Logo

Agilent Technologies 7000B Series - To Trigger on Odd or Even Fields

Agilent Technologies 7000B Series
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggering 4
InfiniiVision 7000B Series Oscilloscopes User’s Guide 147
To trigger on odd or even fields
To check the envelope of your video signals, or to measure worst case
distortion, trigger on the odd or even fields. When Field 1 is selected, the
oscilloscope triggers on color fields 1 or 3. When Field 2 is selected, the
oscilloscope triggers on color fields 2 or 4.
1 Press the [Trigger] key; then, select TV.
2 Press the Settings softkey, then press the Standard softkey to select the
appropriate TV standard.
3 Press the Mode softkey and select Field1 or Field2.
The trigger circuits look for the position of the start of Vertical Sync to
determine the field. But this definition of field does not take into
consideration the phase of the reference subcarrier. When Field 1 is
selected, the trigger system will find any field where the vertical sync

Table of Contents

Related product manuals