EasyManua.ls Logo

Agilent Technologies 7000B Series - To Clean the Oscilloscope; To Perform Self Test

Agilent Technologies 7000B Series
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oscilloscope Utility Settings 18
InfiniiVision 7000B Series Oscilloscopes User’s Guide 401
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
To perform self test
Pressing [Utility] > Service > Start Self Test performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Self Test:
after experiencing abnormal operation.
for additional information to better describe an oscilloscope failure.
to verify proper operation after the oscilloscope has been repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an 80%
confidence level that the oscilloscope is operating properly.
To clean the oscilloscope
1 Remove power from the instrument.
2 Clean the external surfaces of the oscilloscope with a soft cloth
dampened with a mixture of mild detergent and water.
3 Make sure that the instrument is completely dry before reconnecting it
to a power source.

Table of Contents

Related product manuals