Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
To perform self test
Pressing [Utility] > Service > Start Self Test performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Self Test:
• after experiencing abnormal operation.
• for additional information to better describe an oscilloscope failure.
• to verify proper operation after the oscilloscope has been repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an 80%
confidence level that the oscilloscope is operating properly.
To clean the oscilloscope
1 Remove power from the instrument.
2 Clean the external surfaces of the oscilloscope with a soft cloth
dampened with a mixture of mild detergent and water.
3 Make sure that the instrument is completely dry before reconnecting it
to a power source.