288 InfiniiVision 7000B Series Oscilloscopes User’s Guide
I
2
C Serial Decode
To set up the oscilloscope to capture I
2
C signals, see “Setup for I2C
Signals” on page 282.
To set up I
2
C serial decode:
1 Press [Serial]; then, select the I
2
C serial decode mode.
2 Choose 7-bit or 8- bit address size. Use 8- bit address size to include the
R/W bit as part of the address value, or choose 7- bit address size to
exclude the R/W bit from the address value.
3 If the decode line does not appear on the display, press the [Serial] key
to turn it on.
4 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
decode data.
You can use the Zoom horizontal sweep mode for easier navigation of the
acquired data.
NOTE
Because of oscilloscope hardware resource limitations, you cannot decode I
2
C data while
LIN triggering is selected.
NOTE
For I
2
C triggering setup see “I2C Triggering” on page 284.
NOTE
If the setup does not produce a stable trigger, the I
2
C signal may be slow enough that the
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
softkey to set the trigger mode from Auto to Normal.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals