Displaying Waveforms 3
InfiniiVision 7000B Series Oscilloscopes User’s Guide 83
To specify the probe skew
When measuring time intervals in the ns range, small differences in cable
length can affect the measurement. Use Skew to remove cable-delay errors
between any two channels.
1 Probe the same point with both probes.
2 Press one of the probes associated channel key.
3 In the Channel Menu, press Probe.
4 In the Channel Probe Menu, press Skew; then, select the desired skew
value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a
total of 200 ns difference.
The skew setting is not affected by recalling the [Default Setup] or by
pressing [Auto Scale].
To calibrate a probe
See “Calibrating Probes” on page 57.