Triggering 4
InfiniiVision 7000B Series Oscilloscopes User’s Guide 103
Trigger Coupling
1 Press the [Mode/Coupling] key.
2 Press the Coupling softkey, then select DC, AC, or LF Reject coupling.
DC coupling allows DC and AC signals into the trigger path.
AC coupling places a 10 Hz high-pass filter in the trigger path
removing any DC offset voltage from the trigger waveform. The
high- pass filter in the External Trigger input path is 3.5 Hz for all
models. Use AC coupling to get a stable edge trigger when your
waveform has a large DC offset.
LF (low frequency) Reject coupling places a 50 kHz high- pass filter in
series with the trigger waveform. Low frequency reject removes any
unwanted low frequency components from a trigger waveform, such
as power line frequencies, that can interfere with proper triggering.
Use this coupling to get a stable edge trigger when your waveform
has low frequency noise.
TV coupling is normally grayed-out, but is automatically selected
when TV trigger is enabled in the Trigger Menu.
Note that Trigger Coupling is independent of Channel Coupling. To change
Channel Coupling see page 79.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals