16 Analyzing Data
344 InfiniiVision 7000B Series Oscilloscopes User’s Guide
Performing Mask Tests
One way to verify a waveform's compliance to a particular set of
parameters is to use mask testing. A mask defines a region of the
oscilloscope's display in which the waveform must remain in order to
comply with chosen parameters. Compliance to the mask is verified
point- by- point across the display. Mask test operates on displayed analog
channels; it does not operate on channels that are not displayed.
To enable mask test order option LMT at time of oscilloscope purchase, or
order N5455A as a stand- alone item after oscilloscope purchase.
To create a mask from a “golden” waveform (Automask)
A golden waveform meets all chosen parameters, and it is the waveform to
which all others will be compared.
1 Configure the oscilloscope to display the golden waveform.
2 Press the [Analyze] key.
3 Press Features; then, select Mask Test.
4 Press Features again to enable mask testing.
5 Press Automask.
6 In the Automask Menu, press the Source softkey and ensure the analog
channel from which to create the mask is selected.
7 Adjust the mask’s horizontal tolerance (± Y) and vertical tolerance (±
X). These are adjustable in graticule divisions or in absolute units (volts
or seconds), selectable using the Units softkey.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals