Getting Started 1
InfiniiVision 7000B Series Oscilloscopes User’s Guide 29
Power-On the Oscilloscope
Power Requirements
Line voltage, frequency, and power:
~Line 100- 120 Vac, 50/60/400 Hz
100- 240 Vac, 50/60 Hz
120 W max
Ventilation Requirements
The air intake and exhaust areas must be free from obstructions.
Unrestricted air flow is required for proper cooling. Always ensure that
the air intake and exhaust areas are free from obstructions.
The fan draws air in from the sides of the oscilloscope and pushes it out
behind the oscilloscope.
When using the oscilloscope in a bench- top setting, provide at least 2”
clearance at the sides and 4" (100 mm) clearance above and behind the
oscilloscope for proper cooling.
To mount the oscilloscope in a rack
The InfiniiVision 7000B series oscilloscopes can be mounted in an
Electronic Industries Association (EIA) standard 19- inch (487- mm) rack
cabinet.
To mount the oscilloscope in a rack, purchase and install the N2732A rack
mount kit. Instructions are included in the kit. For details search for
N2732A at www.agilent.com.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals