EasyManuals Logo

Agilent Technologies 7000B Series User Manual

Agilent Technologies 7000B Series
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #325 background imageLoading...
Page #325 background image
MIL-STD 1553 Triggering and Serial Decode 14
InfiniiVision 7000B Series Oscilloscopes User’s Guide 325
Setup for MIL-STD 1553 Signals
To set up the oscilloscope for MIL- STD 1553 signals, these sofkeys appear
in both the Trigger Menu and the Serial Decode Menu:
Upper Ch — selects the channel used for the upper threshold. Because
defined pairs of channels are used for upper and lower thresholds
(either channels 1 and 2, or channels 3 and 4), this softkey also selects
the channel for the lower threshhold.
Lower Ch — shows the channel selected for lower thresholds.
Auto Setup — performs these actions:
Copies the Upper Threshold Channel position and volts/div settings
to the Lower Threshold Channel.
Sets both the Upper Threshold Channel and the Lower Threshold
Channel probe attenuation to 10:1.
Sets the Upper Threshold Channel trigger level to 500 mV and the
Lower Threshold Channel trigger level to - 500 mV.
Turns off trigger noise reject.
Turns on Serial Decode.
Sets the trigger type to MIL- 1553.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals