EasyManua.ls Logo

Agilent Technologies 7000B Series - Page 233

Agilent Technologies 7000B Series
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition Modes 7
InfiniiVision 7000B Series Oscilloscopes User’s Guide 233
Whenever both channels in a channel pair are on, the sample rate for all
channels is halved. For example, when channels 1, 2, and 3 are on, the
sample rate for all channels is 2 GSa/s.
When Realtime sampling is on, the bandwidth of the oscilloscope is
limited because the bandwidth of the reconstruction filter is set to f
s
/4.
For example, a 710x oscilloscope with channels 1 and 2 on has a
bandwidth of 500 MHz when Realtime sampling on and 1 GHz when
Realtime sampling is off.
To see the sample rate, press the [Horiz] key on the front panel. The
sample rate is displayed in the line just above the softkeys.
Sample
rate

Table of Contents