D: Model 2450 in a Model 2400 application Model 2450 Interactive SourceMeter® Instrument
D-16 2450-901-01 Rev. B/September 2013
Model 2400 command
[:SENSe[1]]:RESistance:HOLDoff:DELay
[:SENSe[1]]:RESistance:HOLDoff:DELay?
Model 2450 command
Notes
Model 2400 command
[:SENSe[1]]:RESistance:MODE
[:SENSe[1]]:RESistance:MODE?
Model 2450 command
[:SENSe[1]]:<function>:MODE, where <function> is RESistance.
[:SENSe[1]]:<function>:MODE?
Notes
Model 2400 command
[:SENSe[1]]:RESistance:OCOMpensated
[:SENSe[1]]:RESistance:OCOMpensated?
Model 2450 command
[:SENSe[1]]:<function>:OCOMpensated (on page 6-49), where <function> is
RESistance.
[:SENSe[1]]:<function>:OCOMpensated?
Notes
Model 2400 command
[:SENSe[1]]:VOLTage[:DC]:PROTection[:LEVel]
[:SENSe[1]]:VOLTage[:DC]:PROTection[:LEVel]?
Model 2450 command
:SOURce[1]:<function>:PROTection[:LEVel] (on page 6-73), where
<function> is VOLTage[:DC].
:SOURce[1]:VOLTage:ILIMit[:LEVel]?
Notes
Model 2400 command
[:SENSe[1]]:VOLTage[:DC]:PROTection:TRIPped?
Model 2450 command
:SOURce[1]:<function>:PROTection[:LEVel]:TRIPped? (on page 6-74), where
<function> is VOLTage[:DC].
Notes
SOURce[1] subsystem
Model 2400 command
:SOURce[1]:CLEar:AUTO
:SOURce[1]:CLEar:AUTO?
Model 2450 command
Not available
Notes
Use the Trigger model (on page 3-95
).
Model 2400 command
:SOURce[1]:CLEar:AUTO:MODE
:SOURce[1]:CLEar:AUTO:MODE?
Model 2450 command
Notes
Use the Trigger model (on page 3-95).
Model 2400 command
:SOURce[1]:CLEar[:IMMediate]
Model 2450 command
Not available
Notes
Use the Trigger model (on page 3-95).