D: Model 2450 in a Model 2400 application Model 2450 Interactive SourceMeter® Instrument
D-18 2450-901-01 Rev. B/September 2013
Model 2400 command
:SOURce[1]:CURRent:RANGe?
:SOURce[1]:VOLTage:RANGe
:SOURce[1]:VOLTage:RANGe?
Model 2450 command
:SOURce[1]:<function>:RANGe (on page 6-74), where <function> is
CURRent or VOLTage.
:SOURce[1]:<function>:RANGe?
Notes
Model 2400 command
:SOURce[1]:CURRent:RANGe:AUTO
:SOURce[1]:CURRent:RANGe:AUTO?
:SOURce[1]:VOLTage:RANGe:AUTO
:SOURce[1]:VOLTage:RANGe:AUTO?
Model 2450 command
:SOURce[1]:<function>:RANGe:AUTO (on page 6-75
), where
is
CURRent or VOLTage.
:SOURce[1]:<function>:RANGe:AUTO?
Notes
Model 2400 command
:SOURce[1]:CURRent:SPAN?
:SOURce[1]:VOLTage:SPAN
Model 2450 command
:SOURce[1]:SWEep:<function>:LINear (on page 6-81)
:SOURce[1]:SWEep:<function>:LINear:STEP (on page 6-83)
:SOURce[1]:SWEep:<function>:LIST (on page 6-85)
:SOURce[1]:SWEep:<function>:LOG (on page 6-87)
Notes
Sweep parameters are built into the sweep command path. See Sweep
operation (on page 3-51) for information.
Model 2400 command
:SOURce[1]:CURRent:STARt?
:SOURce[1]:VOLTage:STARt
:SOURce[1]:VOLTage:STARt?
Model 2450 command
:SOURce[1]:SWEep:<function>:LINear (on page 6-81)
:SOURce[1]:SWEep:<function>:LINear:STEP (on page 6-83)
:SOURce[1]:SWEep:<function>:LIST (on page 6-85)
:SOURce[1]:SWEep:<function>:LOG (on page 6-87)
Notes
Sweep parameters are built into the sweep command path. See Sweep
operation (on page 3-51) for information.
Model 2400 command
:SOURce[1]:CURRent:STEP?
:SOURce[1]:VOLTage:STEP
Model 2450 command
:SOURce[1]:SWEep:<function>:LINear (on page 6-81)
:SOURce[1]:SWEep:<function>:LINear:STEP (on page 6-83)
:SOURce[1]:SWEep:<function>:LIST (on page 6-85)
:SOURce[1]:SWEep:<function>:LOG (on page 6-87)
Notes
Sweep parameters are built into the sweep command path. See Sweep
operation (on page 3-51) for information.