TOS93 Series User’s Manual 73
Withstanding Voltage and Insulation Resistance Tests | Setting Test Conditions
Current measurement mode (Current RMS)
This can be set only for AC withstanding voltage (ACW) tests.
Sets the current measurement mode. An i
con appears during testing according to the setting.
1
On the Home menu screen, press Measure repeatedly until Measure is selected.
Each time you press the key, the setting changes between Measure and Scale.
2
Press Current RMS, and then turn the rotary knob to change the value.
This completes the setting.
Value Description
Icon
TrueRMS Measures in true rms values.
Average Measures by converting the mean-value responses to rms values.