74 User’s Manual TOS93 Series
Withstanding Voltage and Insulation Resistance Tests | Setting Test Conditions
Voltage measurement mode (Volt Measure)
This can be set only for AC withstanding voltage (ACW) and DC withstanding voltage (DCW) tests.
Sets the voltage measurement mode. An icon appears during testing according to the setting.
1
On the Home menu screen, press Measure repeatedly until Measure is selected.
Each time you press the key, the setting changes between Measure and Scale.
2
Press Volt Measure, and then turn the rotary knob to change the value.
This completes the setting.
Value Description Icon
RMS (ACW only) Measures in true rms values.
Average (DCW only) Measures in average values.
Peak Measures in peak values.
Example: Home menu screen of ACW

Table of Contents