TOS93 Series User’s Manual 83
Withstanding Voltage and Insulation Resistance Tests | Starting a Test
Starting a test
1
Check that the product is correctly connected to the EUT.
2
Check that “READY” is shown on the display, and then press START.
The DANGER LED lights, and the test starts.
If the test does not start, see “Unable to start a test”
(p.336).
If you want to start another test after the test finishes, set a rest time
(p.82) between tests as
necessary.
WARNING
Risk of electric shock.
• While the DANGER L
ED is lit, do not touch the EUT, test leads, test probes, output
terminals, or other peripheral components.
• When performing a test, be sure to wear rubber gloves for electrical work.
Example of ACW