TOS93 Series User’s Manual 95
Earth Continuity Test | Setting Test Conditions
Test time
Set the test time to on or off.
1
On the Home menu screen, press Time > Test Time.
2
Press Test Time to switch between on and off.
Test Time switches between on and off each time you press the key.
3
If you select on, use the numeric keypad or the rotary knob to enter the time.
Setting range: 0.1 s to 1 000.0 s
This completes the setting.
Value Description
ON Set the time from when the Current rise time (p.96) has passed to the end of the test. If upper
limit judgment (Upper-FAIL) or lower limit judgment (Lower-FAIL) does not occur during a
test, the test will result in PASS.
OFF The test continues until the STOP switch is pressed. Pass judgments are not made.