No. 99MBB122A
3-4
Attaching the detector
IMPORTANT
When fitting the detector into the drive unit, do not force the detector. Doing so
may damage the instrument.
The detector moves smoothly at first being fitted with the guide way in the drive
unit, then pins of the connectors on the detector and the drive unit must be fitted
to each other. After the detector is felt tight in the guide of the drive unit, further
push in the detector until it stops with the pins on the connectors fitted.
When the detector (the standard/retracting type) is firmly fitted in the drive unit,
the screw position on the top of the detector is aligned with the end face of the
drive unit, as shown in the following illustration of the standard detector type.
After the detector is attached
1 Reposition the drive unit to the origin point position.
Once you have confirmed that the drive unit is at the origin point position, proceed to
step 2.
TIP
• When the detector is in retracted position, you must first release the retraction. For
information about releasing the detector from retraction, refer to 15.2, “Retracting the
Detector”.
a Press the [POWER/DATA] key to turn on the power.
b Press the [START/STOP] key to move the connector position of the drive unit to
the origin.
When you have moved the drive unit to the origin point position, the drive unit
returns to the origin after performing a measurement.
c Press the [Esc/Guide] key to turn the power off.