27.2.1. Automatic correction of reference mass
It concerns the < Parts counting> working mode
Working mode < Parts counting> comprises a special function
<
Automatic correction of reference mass>, that can be used for
correcting the unit mass <SMP>. To enable the function in parameters you
need to:
• Enter parameter group < Working modes> according to ch.
25 of this manual, choose: „
Parts counting / Automatic
correction of reference mass” and set appropriate option.
Options:
- Automatic correction of reference mass disabled
- Automatic correction of reference mass enabled
Function < Automatic correction of reference mass> in mode
<
Parts counting > is enabled at the moment of estimating the sample
quantity and signalled by displaying <PCS> and <SMP> (single part/piece
mass) on the top part of the display.
There are four criteria of working “Automatic correction of reference
mass”
function:
1. equilibrium should be reached,
2. quantity of parts/pieces should be increased,
3. added quantity of parts/pieces should not be greater than double
number of parts/pieces on the pan,
4. the new sample can be different from the old sample by ± 0.3
of pcs (absolute value),
If a user recognises that the sample quantity is adequate the unit mass (single
piece mass) can be saved (see ch. 27.6 of this manual) and disable the
function by pressing
.