312
Caution:
With enabled <
Automatic adjustment> parameter it is important to
periodically carry out manual internal adjustment.
54.4. Internal adjustment
Internal adjustment is carried out by means of internal adjustment weight.
Procedure:
Enter <
Weighing / Internal adjustment> submenu, the following
message is displayed:
Upon completed adjustment procedure the following message is
displayed:
Press button to confirm.
Caution:
The adjustment has to be carried out when the weighing pan is unloaded and
working conditions are stable (no drafts and vibrations).