80
• Accept the message box by pressing key and return to weighing
mode.
< Setting of start mass> parameter allows to adjust
start mass of platform 1.
Caution:
The factory adjustment process for platforms 2, 3, 4 is analogical to the one
described above.
22.2. Start mass adjustment
It is possi
ble to adjust only a start mass, it helps to correct the start zero
when the span does not change.
Procedure:
• Enter submenu < User Adjustment > according to ch. 22 and select:
“
Setting of start mass”,
• After entering the parameter the following message box appears:
• Take the load off the pan of platform 1,
• Press button . The following message appears during adjusting
start mass: „Evaluation of starting mass”,