Example:
S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value
in basic unit is returned immediately.
40.4.7. Send immediate results from all platforms in basic units
Syntax: SIA CR LF
Possi
ble answers:
SIA_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- mass values are immediately returned from all
platforms in basic units
Frame format with mass from subsequent platforms as indicator reply:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
P n space stability space sign mass space unit CR LF
n - weighing platform number
mass - 9 characters justified to the right
unit - 3 characters justified to the left
Example:
Let us assume that both platforms are connected to scale.
S I A CR LF – computer command
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF - command done, mass values from
both platforms are returned in basic units
40.4.8. Send the stable result in current unit
Syntax: SU CR LF
Possible answers: