3. Barcode scanners can send additional information apart from the expected
barcode e.g. symbology (type of barcode). It is advisable to disable such
information because RADWAG devices and software do not use it.
4. Some RADWAG systems can omit unnecessary information by using
parameters that mark the beginning and the length of the code required to
analyse.
5. A special protocol is required in order the code be received by RADWAG
equipment. It is required to program an appropriate prefix and suffix. Prefix
– one byte 01 hexadecimally, suffix one byte 0D hexadecimally.
6. Most barcode scanners allow to enable/disable different symbologies
(barcode types).
7. Programming of scanners is usually performed by reading
special barcodes or by using an external software tool.
8. Scanners marketed together with RADWAG systems are
always configured according to the rules above.
Barcode with required prefix and
suffix in hexadecimal format
Barcode without
required –fixes in
ASCII format
Code type
01 30 30 32 31 30 31 32 36 0D 00210126 EAN-8
01 30 31 32 33 34 35 36 37 38 39 0D 0123456789 2 of 5
01 43 4F 44 45 20 33 39 20 54 45 53 54 0D CODE 39 TEST CODE 39
01 31 31 30 31 32 33 34 35 36 37 38 39 31 0D 1101234567891 EAN-13
01 43 6F 64 65 20 31 32 38 20 54 65 73 74 0D CODE 128 Test CODE 128
53. APPENDIX 07 – Computer Program „PGC Viewer”
Computer program „PGC Viewer” is intended for viewing and printing reports
from controls and average tare tests.
Caution:
Installation program „PGC Viewer” is recorded on the CD connected to this
product.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PUE 7.1 and is the answer not in the manual?

RADWAG PUE 7.1 Specifications

General IconGeneral
Keypadmembrane
CompatibilityRadwag Balances
HousingPlastic
Operating Temperature0°C to 40°C
InterfacesRS-232

Related product manuals