EasyManua.ls Logo

RADWAG PUE 7.1

RADWAG PUE 7.1
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
310
Autozero
Enter this parameter to enable automatic control and correction of
zero indication. There are, however, some cases when this
function can be a disturbing factor for the measuring process. e.g.
very slow placing of a load on the weighing pan (load adding, e.g.
pouring, filling). In such case, it is recommended to disable the
function. Available options:
- function disabled, - function
enabled.
Last digit
Enter this parameter to enable/disable display of the last digit
(placed on the right of the decimal point) of the weighing result.
Available values: Always: all digits are displayed; Never: last digit
disabled; When stable: last digit displayed only when the result is
stable.
54.2. LO Threshold for automatic functions
<LO threshold> parameter allows you to configure the function of automatic
operation. The next measurement is saved when mass indication gets below
the set net value of <LO threshold> parameter.
Procedure:
Enter <
Weighing / LO threshold> submenu, an on-screen
keyboard is displayed.
Enter LO threshold value and press
button.
54.3. Automatic adjustment
Automatic adjustment using internal adjustment weight is carried out in 3
different cases:
Triggered by plugging the instrument to the mains.
Triggered by temperature change (by 3°C).
Triggered by passage of specified time interval (every 3h).
Enabling <
Automatic adjustment> parameter:
Enter <
Weighing> submenu.
Select <
Automatic adjustment> parameter and set respective value
(
- function disabled, - function enabled).
Automatic adjustment process:
Upon detecting the need for carrying out an adjustment process, the
following message is displayed: