54.5. Automatic adjustment time-triggered
The function enables carrying out automatic adjustment each day at a specific
day and time. You can declare up to 5 values (hours) per day (e.g. 03.00, 8.00,
13.00, 18.00, 23.00).
Procedure:
Enter<
Weighing / Automatic adjustment time-triggered>
submenu and select <
Automatic adjustment – set time (h:m)>
parameter.
Enter <
Date> parameter and enter respective hour of automatic
adjustment start (in the following order: hour, minute).
Entered value is automatically set as the time of internal adjustment start.
To disable the function, uncheck <
Active> parameter ( - function
disabled,
- function enabled).