40.4.20. Give balance serial number
Syntax: NB CR LF
Possible answers:
NB_A_”x” CR LF
- command understood, returns serial number
NB_I CR LF
- command understood but not accessible at this moment
x – serial number of the device (inserted in between inverted commas)
Example:
NB CR LF – give serial number
NB_A_”123456” CR LF – serial number of the device – 123456
40.4.21. Set mass value of a single item
Syntax: SM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Possible answers:
SM_OK CR LF
- command understood
SM_I CR LF
- command understood but not accessible at this moment
(e.g. mode other than „Parts counting”)
ES CR LF
- command not recognised (mass format incorrect)
Caution:
Command supported for „Parts counting” mode exclusively.
40.4.22. Set reference mass value
Syntax: RM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Possible answers:
RM_OK CR LF
- command understood
RM_I CR LF
- command understood but not accessible at this moment
(e.g. mode other than „Deviations”)
ES CR LF
- command not recognised (mass format incorrect)
Caution:
Command supported for „Deviations” mode exclusively.