Working modes list:
1 – Weighing
2 – Parts Counting
3 – Deviations
4 – Dosing
5 – Recipes
6 –Weighing Animals
7 – Density
17 - Control of packaged goods PGC
20 - Comparator
21 – Vehicle scale
Caution:
Working modes numbering is identical for each kind of balance. The numbers
are assigned to working modes names. Some balances give only the number
as a response.
Example:
OMI CR LF - give accessible working modes
OMI CR LF - accessible working modes are given
2_”Parts Counting” CR LF - number + name
4_”Dosing” CR LF
5_“Recipes” CR LF
OK CR LF - command carried out
40.4.25. Set working mode
Syntax: OMS_n CR LF
Where: n – parameter, decimal value determining working mode number.
To see detailed description go to OMI command.
Possible answers:
OMS_OK CR LF
- command carried out
OMS_I CR LF
- command understood but not accessible at this moment
OMS_E CR LF
- error in-course of command execution, no parameter
or incorrect format