109
28.3. Reference unit mass estimated by weighment
Procedur
e:
• Enter < Deviations> according to ch. 28.1 of this manual,
• If the standard is to be weighed in a container, the container needs to be
put on the pan and tarred,
• Press
(Estimate standard mass), then a message is displayed:
<Put standard>,
• Put the load on the pan. After stabilization the result is taken as a standard
(symbol
). Confirm it by pressing ,
• At the same time the weight unit is changed to (w %).
28.4. Reference unit mass inscribing into the memory
Procedur
e:
• Enter < Deviations> according to ch. 28.1 of this manual,
• Press
(Give piece mass), then an editing field is displayed <Give
piece mass> together with the screen keyboard,
• Enter a value and confirm by pressing
,
• At the same time the weight unit is changed to (w %).
29. WORKING MODE - DOSING
< Dosing> mode enables carrying out filling and dosing processes of goods
on scales featuring a indicator PUE 7.1 series. The working mode enables
manual and automatic dosing on a single or simultaneously on many weighing
platforms.