264
40.5. Manual printouts / automatic printouts
Users ca
n general manual or automatic printouts from the scale.
Manual printouts can be performed after loading the pan
and stabilizing indication by pressing
.
Automatic printouts can be performed only after loading
the pan and stabilizing indication.
Format frame:
1 2 3 4 -12 13 14 15 16 17 18
stability space sign mass space unit CR LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
sign
[space] for positive values or
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – the printout generated from
the scale after pressing
.
41. MODBUS RTU COMMUNICATION PROTOCOL
The indicator PUE 7.1 enables complete operation of programmable controllers
(PLC), which communicate through Modbus protocol class 0.
The indicator in a network using modbus standard operates as slave on
an accessible interfaces, RS232 or Ethernet. According to the standard,
Modbus class 0, the device uses two functions:
Read multiple registers (fc 3), for cyclic readout of signals from
the weighing module,