139
Caution:
If before control start the operator:
• Does not remove a load from the weighing pan or other criteria for zeroing
are failed (e.g. unstable weighing result) the scale displays a message:
<Unable to start control. Zeroing error>,
• Does not log in or the logged operator is not authorized to carry out
the control, the scale displays a message: <No authorization>,
• Does not select a product from the database, the scale displays a
message: <Product not selected >,
• Does not declare batch quantity, the scale displays a message: <Batch
quantity not entered >.
31.6. Control aborting procedure
After control start an o
perator can abort the control process in an optional
moment by pressing a function hot-key
(control stop) located in the
bottom bar of the control process window.
Pressing the (control stop) key is followed by displaying a message:
Pressing key causes reuniting to the control process in progress.
Pressing
key causes terminating the control and returning to settings of
the PGC mode.
Simultaneously, a control report with status <Terminated> is added to the
database of <
Controls>.