78
10 [s]
- Displaying an information about errors for 10 [s]
Max
- Displaying an information about errors until it has been
confirmed with
button.
21.15. Export / import of settings
The user
can export/import scale settings (printout templates, user parameters)
using flash memory stick.
Procedure for settings export:
Connect flash memory stick to USB port of your scale,
Enter
Others / Export” submenu, settings are exported
automatically to flash memory stick plugged to scale USB port.
Upon completed procedure the following message is displayed:
„Operation completed successfully”.
Procedure for settings import:
Connect flash memory stick to USB port of your scale,
Enter
Others / Import” submenu, settings are imported
automatically from flash memory stick plugged to scale USB port.
Upon completed procedure the following message is displayed:
„Operation completed successfully”.
22. USER ADJUSTMENT
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g.
g, kg etc.). In order to do this they require an adjustment factor. It is adjusted
during the adjustment procedure using a mass standard. Adjustment should
be carried out if weighing a standard mass shows a different mass value.
To enter < User Adjustment>, press key and then:
User Adjustment”.