In order to specify number of weighings that are to be subjected to deletion
process it is necessary to activate option: <
Ask for number of weighings
to be deleted>.
Procedure:
Enter submenu: Reports / Reports configuration”
according to ch. 38 of this manual,
Activate option <
Ask for number of weighings to be deleted>,
Upon returning to home screen and pressing programmable button
< Undo weighing>, message: <Number of weighings to be
deleted> with a numeric keyboard is displayed,
Enter appropriate value (from 1 to 10) and press button
for confirmation, message: <Number of deleted weighings: x>
is displayed, where x – number of deleted weighings.
Caution:
When:
Some of the weighings that are to be deleted refer to any kind of report
(report on dosing, report on formula, report on PGC control, report on SQC
control, report on vehicle scale, report on density, report on differential
weighing) then the following message is displayed <Number of deleted
weighings: x. Remaining weighings are report-related>.
All of the weighings that are to be deleted refer to any kind of report (report
on dosing, report on formula, report on PGC control, report on SQC
control, report on vehicle scale, report on density, report on differential
weighing) then the following message is displayed <Operation not
allowed. Weighings are report-related>.