The scale does not report an error, the measurement is stable
in the second measuring range.
• LO – the value of LO threshold in calibration/adjustment unit is returned.
• Process status – determines the status of dosing process
byte no.
Decimal value of
a variable (Dec)
process status
B1 B0
0
process disabled (inactive) 0 0
1
dosing start 0 1
2
dosing stop 1 0
3
end of dosing process 1 1
• Inputs status – determines the status of inputs in a indicator PUE7:
Input no. 12 11 10 9 8 7 6 5 4 3 2 1
Decimal value of
a variable (Dec)
OFF 0 0 0 0 0 0 0 0 0 0 0 0 0
ON 1 1 1 1 1 1 1 1 1 1 1 1 4094
• MIN – set value of MIN Threshold (in calibration/adjustment unit)
is returned.
• MAX – set value of MAX Threshold (in calibration/adjustment unit)
is returned.
• Lot number – response is the value of lot number (numeric range
0 - 2147483647).
• Operator – response is the value of a logged operator (numeric range
0 - 32767).
• Product – response is the value of a selected product (numeric range
0 - 32767).
• Customer – response is the value of a code of a selected customer
(numeric range 0 - 32767).
• Packaging – response is the value of a code of a selected packaging
(numeric range 0 - 32767).