272
command – basic command. Setting of an appropriate command’s
byte is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task Dec
0 scale zeroing 1
1 scale tarring 2
5 dosing process start 32
6 dosing process stop (emergency stop) 64
For instance:
0000 0000 0010 0000 – command carried out, dosing process start
complex command - Setting of an appropriate command’s byte
is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task Dec
0 setting tare value 1
1
setting LO threshold value
2
2 setting outputs status 4
3
setting MIN threshold value
8
4
setting MAX threshold value
16
For instance:
0000 0000 0000 0010 – command carried out, LO threshold set to value
given in LO parameter (address 10); See Table “Input variables”.
Caution:
A complex command requires setting an appropriate address parameter
from 6 to 28. See table “Input variables”.
Platform – complex command parameter: number of a weighing platform.
tare – complex command parameter – tare value (in calibration /
adjustment unit).