EasyManua.ls Logo

Waters Autospec Premier - Page 66

Waters Autospec Premier
354 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-2 Scanning techniques
MS method editor with a single function
You can save a function list in a MassLynx experiment file, and then reopen
and edit it at any time.
See “The MassLynx experiment file” on page 3-4.
The figure “MS method editor with a single function” on page 3-2 shows a
simple Function List containing only one function: a Magnet scan between 50
and 500 Da, using EI+ ionization. Immediately above the function bar display
is a time scale, which shows the elapsed time at which the function becomes
active and for how long it runs. In this case, the function starts after
5 minutes and then runs for 30 minutes, terminating after a total elapsed
time of 35 minutes.
Multiple function acquisitions are typically used to acquire different VSIR
groups over different retention windows. A Function List, with two VSIR
functions, each running sequentially for 10 minutes, is shown in figure “MS
method editor with two functions:” on page 3-2.
Tip: Do not start the first function at 0 minutes if it takes some time for the
sample to elute; set the start time to account for this. This initial wait is called
solvent delay, because the solvent usually elutes several minutes after
injection.
MS method editor with two functions:

Table of Contents

Related product manuals