Agilent 4155C/4156C SCPI Command Reference, Edition 1 6-5
Error Messages
Command Error
Error -105 GET not allowed
A group execute trigger was received within a program message.
Error -108 Parameter not allowed
Too many parameters for the command were received.
Error -109 Missing parameter
Fewer parameters were received than required for the command.
Error -110 Command header error
An error was detected in the header. This error message is reported if the
4155C/4156C cannot determine the more specific header errors -111 through -114.
Error -111 Header separator error
An illegal character for a header separator was received; for example, no white
space between the header and parameter.
Error -112 Program mnemonic too long
A keyword in the command header contains more than twelve characters.
Error -113 Undefined header
An undefined command header was received; for example, *XYZ.
Error -114 Header suffix out of range
The value of a numeric suffix attached to a program mnemonic is out of range; for
example, :PAGE:CHAN:SMU7:MODE V specifies illegal SMU number 7.
Error -120 Numeric data error
Numeric (including the nondecimal numeric types) data error. This error message is
reported when the 4155C/4156C cannot determine the more specific errors -121
through -128.
Error -121 Invalid character in number
An invalid character for the data type was received; for example, an alphacharacter
was received when the type was decimal numeric.