Measurements and Math Functions 5
InfiniiVision 7000B Series Oscilloscopes User’s Guide 203
For the best vertical accuracy on peak measurements:
Make sure the probe attenuation is set correctly. The probe attenuation
is set from the Channel Menu if the operand is a channel.
Set the source sensitivity so that the input signal is near full screen,
but not clipped.
Use the Flat Top window.
Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
Use the Hanning window.
Use Cursors to place an X cursor on the frequency of interest.
Adjust frequency span for better cursor placement.
Return to the Cursors Menu to fine tune the X cursor.
For more information on the use of FFTs please refer to Agilent
Application Note 243, ”The Fundamentals of Signal Analysis” at
http://cp.literature.agilent.com/litweb/pdf/5952- 8898E.pdf. Additional
information can be obtained from Chapter 4 of the book “Spectrum and
Network Measurements” by Robert A. Witte.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000B Series and is the answer not in the manual?

Agilent Technologies 7000B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals