7 Acquisition Modes
226 InfiniiVision 7000B Series Oscilloscopes User’s Guide
When the oscilloscope triggers, the single acquisition is displayed and
the oscilloscope is stopped (the [Run/Stop] key is illuminated in red).
4 To acquire another waveform, press [Single] again.
Auto Single
Auto- trigger will generate a trigger for you if one is not found in the
predetermined time (about 40 ms) after you press [Single]. If you want to
make a single acquisition and you are not particularly interested in
triggering the acquisition (for example, if you are probing a DC level), set
the trigger mode to Auto (see page 99) and press the [Single] key. If a
trigger condition occurs, it will be used; if a trigger doesn’t happen, an
untriggered acquisition will be taken.