EasyManua.ls Logo

Keysight U8903B - Output Configuration (Analog Generator); Figure 3-56 Analog Generator > Output Config Menu Page; Table 3-28 Analog Generator > Output Config Menu Description

Keysight U8903B
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Audio Generator Functions 3
Keysight U8903B User’s Guide 141
Output Configuration (Analog Generator)
The output configuration for the analog generator is displayed as shown in Figure 3-56.
Figure 3-56 Analog Generator > Output Config menu page
Table 3-28 Analog Generator > Output Config menu description
Menu Description
Connector
Press the Connector softkey to select the output connector type.
–Bal
Balanced mode outputs a pair of differential signals which are equal in amplitude but 180 degrees out of phase on the
XLR positive and negative pins.
–UnBal
Unbalanced mode outputs a signal referenced to ground on the BNC output connector.
–Com
Common mode outputs a pair of equal amplitude and in-phase signals on the XLR positive and negative pins. The
common mode test signal is applied to both pins 2 and 3 or the XLR connector, while pin 1 is connected to the
return signal.
IEC60268
The common mode test signal is applied to both pins 2 and 3 or the XLR connector, while pin 1 is connected to the
return signal. An additional 10 Ω output series resistance is added to pin 2 or 3.
Impedance
Press the Impedance softkey to select the output impedance value.
For Bal, Com, and IEC60268
600 Ω
100 Ω
–40 Ω
For UnBal
600 Ω
–50 Ω
–20 Ω

Table of Contents