Test Sequence Application 7
Keysight U8903B User’s Guide 321
Signal analysis
The phase signal analysis settings menu page is displayed as shown in Figure 7-22.
Figure 7-22 Phase > Signal Analysis settings menu page
Table 7-21 Phase > Signal Analysis settings menu description
Menu Description
Ref Channel
Press the Ref Channel softkey to set the reference channel number. The phase of each channel is measured against the
reference channel. The phase result for the reference channel should always be displayed zero.
Sample Size
Press the Sample Size softkey to select the number of samples to be acquired for the measurement.
2048
4096
8192
16384
32768
65536
131072
262144
524288
–1M
–2M
Input Range
Press the Input Range softkey to configure the input range settings.
Track 1st Ch
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to
channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual
channels.
Voltage Range
Set the input voltage range.