Audio Analyzer Measurement Functions 4
Keysight U8903B User’s Guide 245
Figure 4-61 Bluetooth Analyzer > Meas Config > Trigger menu page
Table 4-30 Digital Analyzer > Meas Config > Trigger menu description
Menu Description
Trigger Source
Press the Trigger Source softkey to select the trigger source. The default trigger in is Free Run, where the analyzer triggers
immediately after the previous data is acquired. If External trigger in is set, the analyzer waits for a trigger pulse on the
Trigger In connector at the rear panel before acquiring the measurement data.
–Free Run
–External
– Channel
Trigger Edge
This is only available when the trigger source is External or Channel.
Press the Trigger Edge softkey to select the trigger edge type.
– Rising
–Falling
Interface
This is only available when the trigger source is Channel.
Press the Interface softkey to select the trigger interface.
– Analog
– Digital
–Bluetooth
Channel
This is only available when the trigger source is Channel.
Press the Channel softkey to set the channel number.
Trigger Level
This is only available when the trigger source is Channel.
When the Trigger Interface is Analog, the Trigger Level is in Vpeak.
When the Trigger Interface is Digital or Bluetooth, the Trigger Level is in Hex.
Press the Trigger Level softkey to set the trigger level.