Sweep Function 6
Keysight U8903B User’s Guide 283
Points settings
Figure 6-3 Sweep Parameter > Points Settings menu page 1
Table 6-2 Sweep Parameter > Points Settings menu description
Menu Description
Spacing
Press the Spacing softkey to select the spacing type.
Log – Geometric series
Linear – Arithmetic series
Custom – Arbitrary spacing
Unit
Press the Unit softkey to select the sweep unit. This setting is only available when the sweep parameter is amplitude.
–Vrms
–dBV
–Vp
–Vpp
–dBm
–dBu
–dBSPL
Start Press the Start softkey to set the sweep start value. This setting is only available when the sweep spacing is Log or Linear.
Stop Press the Stop softkey to set the sweep stop value. This setting is only available when the sweep spacing is Log or Linear.
Step
Press the Step softkey to set the sweep step value. Adjusting the sweep step value will change the sweep points value.
This setting is only available when the sweep spacing is Log or Linear.
Step size 10
10
Stop
Start
---------
⎝⎠
⎛⎞
log
Points 1()
----------- --------------
=
Step size
Start Stop()
Points 1()
----------- ------------ ----=