7 Test Sequence Application
364 Keysight U8903B User’s Guide
Table 7-44 Stepped Level Sweep > Signal Generation settings menu description
Menu Description
Waveform
Press the Waveform softkey to select the waveform type.
–Sine
– Arbitrary
Output Press the Output softkey to select the output channel.
Sweep Config
Press the Sweep Config softkey to set the sweep configuration.
–Start
Set the start value for the sweep parameter.
–Stop
Set the stop value for the sweep parameter.
–Spacing
Select Log, Linear, or Custom for the sweep spacing.
–Points
Set the number of sweep points.
–Step Size
Set the step size for linear spacing. This setting is only available when spacing is set to Linear.
– Dwell Time
Set the delay between the sweep points.
– Edit Points
Edit the individual points value, insert or remove points, load points, and save the points.
Waveform Config
Press the Waveform Config softkey to set the waveform configuration.
– Track 1st Ch
If Track 1st Ch is enabled, the generator channel 1 waveform configuration settings will be copied to the other channels
and the waveform configuration settings for the other channels cannot be edited. Any changes made to channel 1
waveform configuration settings will be reproduced in the other channels waveform configuration settings. Disable the
Track 1st Ch to set the individual channels.
– Frequency
Set the frequency value.
–DC Offset
Set the DC offset value.