HP8903B 8
Keysight U8903B User’s Guide 413
Sweep
In HP8903B mode, the source frequency is logarithmically swept. The number of frequency
points in a sweep is determined by the sweep width (ratio of the stop and start frequencies)
and the selected sweep resolution. The maximum number of points allowable in a sweep is
255. The frequency points in a sweep can be computed by using the following formulas:
Where n is the frequency point number (0 is the start frequency) and k is the number of
points per decade. Refer to the HP8903B special function codes, 17.0 to 17.9 as listed in
Table 8-6. The frequency point formula for reverse sweep is as follows:
Select Sweep in the HP8903B menu page to configure the HP8903B sweep settings.
Figure 8-4 HP8903B > Sweep menu page
Frequency Start frequency 10
n
k
--
⎝⎠
⎛⎞
×=
Frequency Start frequency 10
n
k
----
⎝⎠
⎛⎞
×=
Table 8-5 HP8903B > Sweep menu description
Menu Description
Freq. Start Press the Freq. Start softkey to set the HP8903B sweep frequency start value.
Freq. Stop Press the Freq. Stop softkey to set the HP8903B sweep frequency stop value.
Sweep Status Press the Sweep Status softkey to start or abort sweeping in the HP8903B mode.