Appendixes A
Keysight U8903B User’s Guide 567
Configuration 2
In this configuration, the DUT uses an external master clock as the reference clock. The
U8903B receives the data, frame sync, and bit clock from the DUT. A typical application for
this configuration is analog to digital converter (ADC) evaluation.
Press on the FUNCTION panel to switch between audio generator or audio analyzer
mode, and press on the FUNCTION panel to switch to digital interface.
Figure A-5 DSI test configuration 2
1 At the digital analyzer, press Input Config > Connector, and select DSI as the digital
analyzer input type.
2 At the digital generator, press DSI Config > Master Clock, and select Off to turn off the
master clock.
3 At the digital analyzer, press DSI Config > W/Bclk Dir, and select In as the word and bit
clock direction.
U8903B
DUT
External master clock
Pin 9
Pin 10
Pin 11Bit clock
Frame Sync
Data